MUD TESTING EQUIPMENT
For years, Inmotion Energy has supplied
precision instruments and drilling tools for
testing drilling fluids, well cements, and
completion fluids.
From our Bangalore facility, we provide
customers across India and the Middle East
with quality products and exceptional
service.
INDUSTRIES WE SERVE
— Onshore & offshore drilling contractors
— National & independent oil companies
— Mud engineering & testing laboratories
— Well completion & workover operations
WHY CHOOSE US
— Rigorous quality control on every unit
— Fast turnaround on spares &
replacements
— Field-tested designs, built for uptime
INMOTION ENERGY (OPC) PRIVATE LIMITED
- 186, 1st Main Road, Vijaya Bank Layout, Bilekahalli, Bangalore, Karnataka – 560076
- Phone: +91 96637 44619
- info@inmotionenergy.in
- imotion.in
PRODUCT SPECIFICATION — HTHP Dynamic Linear Swell Meter
Model: IM-LSM-HTHP Compliance / target standard: Advanced R&D Dynamic Swelling
HTHP Dynamic Linear Swell Meter
Figure 12. Reference photograph of the HTHP Dynamic Linear Swell Meter. Components are listed on page 2.
Intended applications
• Shale inhibition testing at bottom-hole temperature and pressure.
• Evaluating oil-based and synthetic-based fluids against reactive shales.
• Wellbore-stability studies with dynamic fluid circulation.
• Research on new inhibitors and additives under realistic conditions.
FEATURES
• Heated pressure vessel with PID temperature control.
• LVDT measurement through a pressure-sealed cap.
• Pressure supply with gauge and relief protection.
• Programmable test temperature and logging of swell versus time.
• Optional fluid circulation for dynamic exposure.
TECHNICAL SPECIFICATIONS
| Sensor | LVDT displacement transducer |
| Sample | Compacted wafer, typically about 25 mm diameter |
| Temperature | Elevated, typically up to about 150-200 C (confirm on datasheet) |
| Pressure | Elevated, typically up to about 1,000-3,000 psi (confirm on datasheet) |
| Fluid mode | Static or dynamic |
| Measured value | Linear swell, % of initial height |
| Control | PID temperature control with data logging |
Main components (see Figure 12)
| No. | Component |
| 1 | LVDT displacement sensor |
| 2 | Pressure gauge |
| 3 | Vessel cap and pressure seal |
| 4 | Controller and display |
| 5 | Heating jacket |
| 6 | Wafer and test fluid |
Principle of operation
Basic operating procedure
1. Prepare and compact the wafer, and place it in the holder.
2. Position the holder and rod in the vessel, fill with test fluid and close the vessel.
3. Connect the LVDT and zero the reading.
4. Pressurise to the test pressure and heat to the test temperature using the controller.
5. Log the swelling continuously for the specified time.
6. Cool below 50 C, vent pressure, open the vessel, clean all parts and report swell versus time.
Calibration and maintenance
• Replace seals and O-rings regularly; use high-temperature lubricant on threads.
• Calibrate the LVDT, thermocouple and pressure gauge annually.
• Inspect the relief device and pressure lines.
• Clean the vessel and rod after each test.
Safety precautions
• Never open the vessel while hot or pressurised.
• Use heat-resistant gloves and a face shield.
• Operate only within rated limits; trained personnel only.
Reference sources
• US Patent 8,434,355 B1, High pressure high temperature linear swell meter
• US Patents 5,275,063 and 6,247,358 (shale hydration and reactivity testing)
• ISO/NP 26652 (under development); OFITE dynamic linear swell meter literature